Malfunction Servicing of Philips T5-NT MRI Set
-
Graphical Abstract
-
Abstract
Half black screen was shown in Philips T5-NT MRI Scanner and with its message. This trouble was associated with environment temperature, checking machine and find the bad parts TC-420 board at the best power up time. The malfunction disappeared after this board was replaced.
-
-