LIU W N, SUN X L, DUAN Y L, et al. Application Value of Dual-Source CT Noise-Optimized Virtual Monoenergetic Imaging Technology in Portal Vein Imaging of Patients with Portal Hypertension[J]. CT Theory and Applications, xxxx, x(x): 1-8. DOI: 10.15953/j.ctta.2024.084. (in Chinese).
Citation: LIU W N, SUN X L, DUAN Y L, et al. Application Value of Dual-Source CT Noise-Optimized Virtual Monoenergetic Imaging Technology in Portal Vein Imaging of Patients with Portal Hypertension[J]. CT Theory and Applications, xxxx, x(x): 1-8. DOI: 10.15953/j.ctta.2024.084. (in Chinese).

Application Value of Dual-Source CT Noise-Optimized Virtual Monoenergetic Imaging Technology in Portal Vein Imaging of Patients with Portal Hypertension

More Information
  • Received Date: June 04, 2024
  • Revised Date: September 17, 2024
  • Accepted Date: September 24, 2024
  • Available Online: November 04, 2024
  • Objective: This study aimed to compare and analyze the effects of different keV values of dual-source computed tomography (CT) noise-optimized virtual monoenergetic imaging (VMI+) on the quality of portal vein imaging in patients with portal hypertension and to determine the optimal keV value for VMI+. Methods: Postprocessing was performed on portal venous phase images of 37 patients with portal hypertension who underwent dual-source CT. The VMI+ technique was used to reconstruct nine groups of monoenergetic images at 40, 45, 50, 55, 60, 65, 70, 75, and 80 keV. Compared with the standard mixed images automatically generated with a linear blending coefficient of 0.6 (M-0.6), the CT values and standard deviation of the main portal vein (MPV) and liver parenchyma were measured for each group of images. The signal-to-noise ratio (SNR) of the MPV and the contrast-to-noise ratio (CNR) with the liver parenchyma were calculated. On the maximum intensity projection images of the portal venous phase, the luminal diameters of the MPV, splenic vein (SPV), left branch of the portal vein (IHLPV), and right branch of the portal vein (IHRPV) were measured, and the number of detected cases of diseased vessels in each image group was recorded. Objective evaluation was conducted using analysis of variance. Subjective evaluation was performed by two deputy chief radiologists who used a five-point scale to assess the contrast effect of the MPV against the background liver parenchyma and the overall image quality in the 10 groups of images. Results: The CT value, SNR, CNR of the MPV, and subjective image scores measured for the 40 ~ 65 keV portal vein groups were higher than those for the M-0.6 group, and the difference was statistically significant. The CT value, SNR, CNR of the MPV, and subjective image scores measured for 70 keV images were similar to those of the M-0.6 group, with no statistical difference. The CT value, SNR, CNR of the MPV, and subjective image scores measured for the 75 and 80 keV groups were lower than those for the M-0.6 group, and the difference was statistically significant. The diameters of the MPV, IHLPV, IHRPV, and SPV were the same among all the groups, and the same number of pathological vessels were detected, with no statistical significance. Among them, the 40 keV group had the highest CT values, SNR, and CNR of the MPV among all 10 groups. Conclusion: The application of VMI+ technology (40 ~ 65 keV) in dual-source CT significantly improves the portal-vein image quality of patients with portal hypertension, which is beneficial for displaying portal vein vessels. The optimal monoenergetic energy level is 40 keV.

  • [1]
    刘腾, 卢俊宇, 何松. 肝硬化门静脉高压诊断方法的研究进展[J]. 重庆医学, 2023, 52(21): 3303−3307. DOI: 10.3969/j.issn.1671-8348.2023.21.019.

    LIU T, LU J Y, H S. Research advances in diagnostic methods of liver cirrhosis portal hypertension[J]. Chongqing Medical Journal, 2023, 52(21): 3303−3307. DOI: 10.3969/j.issn.1671-8348.2023.21.019. (in Chinese).
    [2]
    LAURA T, GUADALUPE G T. Portal Hypertension: Pathogenesis and Diagnosis[J]. Science Direct., 2019, 23(4): 573−587.
    [3]
    高德军, 罗鹰, 陈首名, 等. 双源CT门静脉成像在肝硬化门静脉侧支循环的临床应用[J]. 中国CT和MRI杂志, 2019, 17(8): 100−102,106. DOI: 10.3969/j.issn.1672-5131.2019.08.030.

    GAO D J, Luo Y, CHEN S M, et a1. Clinical application of Dual source CT portographv in portal collateral circulation in hepatocirrhosis[J]. Chinese Journal of CT and MRI, 2019, 17(8): 100−102,106. DOI: 10.3969/j.issn.1672-5131.2019.08.030. (in Chinese).
    [4]
    RENGO M , BELLINI D , De CECCO C N , et al . The optimal contrast media policy in CT of the liver. Part I: Technical notes[J]. Acta Radiologica, 2011, 52(5): 467−472
    [5]
    陈琳钰, 方姝, 陈勇, 等. 虚拟单能图像与常规CT图像的图像质量对比[J]. CT理论与应用研究, 2022, 31(2): 219−226.

    CHEN L Y, FANG S, CHEN Y, et al. Comparison of image quality between virtual monochromatic images and conventional CT images[J]. CT Theory and Applications, 2022, 31(2): 219−226. (in Chinese).
    [6]
    胡斌, 陈静静, 徐文坚, 等. 双能CT改进的虚拟单能成像重建算法的门静脉图像质量评估[J]. 首都医科大学学报, 2019, 40(6): 943−947. DOI: 10.3969/j.issn.1006-7795.2019.06.023.

    HU B, CHEN J J, XU W J, et al. Assessent of portal vein based on a novel virtual monoenergetic algorithm in dual-energy CT[J]. Journal of Capital Medical University, 2019, 40(6): 943−947. DOI: 10.3969/j.issn.1006-7795.2019.06.023. (in Chinese).
    [7]
    ZENG Y W, GENG D Y, ZHANG J. Noise-optimized virtual monoenergetic imaging technology of the third-generation dual-source computed tomography and its clinical applications[J]. Quantitative Imaging in Medicine and Surgery, 2021, 11(11): 4627−4643. DOI: 10.21037/qims-20-1196.
    [8]
    杨桐立, 王照谦, 杨志强, 等. 双能量CT Mono+技术临床应用的研究进展[J]. 临床放射学杂志, 2023, 42(2): 340−344.
    [9]
    周彦宇, 林雁玲. 双源CT虚拟单能量成像技术不同keV值对门静脉成像质量提高的研究[J]. 现代医用影像学, 2022, 31(2): 199−203. DOI: 10.3969/j.issn.1006-7035.2022.2.xdyyyxx202202001.

    ZHOU Y Y, LIN Y L. Study on the improvement of portal vein imaging quality with different keV values of Dual-energy CT virtual single energy imaging technology[J]. Modern Medical Imageology, 2022, 31(2): 199−203. DOI: 10.3969/j.issn.1006-7035.2022.2.xdyyyxx202202001. (in Chinese).
    [10]
    黄亚博, 杨莹, 王晋君. CT门静脉成像参数评估乙型肝炎肝硬化患者肝功能分级及EVB风险的价值[J]. 肝脏, 2024, 29(2): 162−165, 173.

    HUANG Y B, YANG Y, WANG J J. The value of portal vein imaging parameters of computer tomography in evaluating liver function grading and esophageal variceal bleeding risk in hepatitis B patients with cirrhosis[J]. Chinese Hepatology, 2024, 29(2): 162−165, 173. (in Chinese).
    [11]
    莫彩建, 翁建辉, 陈金水, 等. MSCT 联合 MSCTP 技术在肝硬化门静脉高压症术前诊断中的应用价值[J]. 海南医学, 2022, 33(1): 86−89.

    MO C J, WONG J H, CHEN J S, et al. Application value of multi-slice spiral CT combined with multi-slice spiral CT portography in preoperative diagnosis of cirrhotic portal hypertension[J]. Hainan Medical Journal, 2022, 33(1): 86−89. (in Chinese).
    [12]
    杨克柽, 陈伟. 肝脏门静脉期CT增强的影响因素与对策[J]. 国际医学放射学杂志, 2015, 38(6): 562−566,571. DOI: 10.3874/j.issn.1674-1897.2015.06.Z0612.

    YANG K G, CHEN W. Contrast enhancement of abdominal CT at portal venous phase: Considerations and solutions[J]. International Journal of Medical Radiology, 2015, 38(6): 562−566,571. DOI: 10.3874/j.issn.1674-1897.2015.06.Z0612. (in Chinese).
    [13]
    PARAKH A, MACRI F, SAHANI D. Dual-energy computed tomography: Dose reduction, series reduction, and contrast load reduction in dual-energy computed tomography[J]. Radiologic Clinics of North America, 2018, 56: 601−624. DOI: 10.1016/j.rcl.2018.03.002.
    [14]
    王敏, 郭美玲, 李剑. CT低管电压对胸痛三联征排查 图像质量及辐射剂量的影响[J]. 中国医学影像学杂志, 2013, 21(9): 715−717,719. DOI: 10.3969/j.issn.1005-5185.2013.09.023.

    WANG M, GUO M L, LI J. Influence of CT low tube voltage on image quality and radiation dose in patients with chest pain[J]. Chinese Journal of Medical Imaging, 2013, 21(9): 715−717,719. DOI: 10.3969/j.issn.1005-5185.2013.09.023. (in Chinese).
    [15]
    D’HGELO T, CICEM G, MZIOTTI S, et a1. Dual energy computed tomography virual monoenergetic imaging: Technique clinical applications[J]. British Journal of Radiology, 2019, 92: 20180546.
    [16]
    王珍, 张声旺, 顾潜彪, 等. CT影像组学评估门静脉高压及预测曲张静脉出血[J]. 中国医学影像技术, 2020, 36(12): 1896−1899.

    WANG Z, ZHANG S W, GU Q B, et al. CT radiomics for assessing portal hypertension and predicting variceal hemorrhage[J]. Chinese Journal of Medical Imaging Technology, 2020, 36(12): 1896−1899. (in Chinese).
    [17]
    崔楚坤, 李家欢, 曾鹏程, 等. 多层螺旋CT血管造影在冠状动脉病变程度诊断中的临床价值[J]. 海南医学, 2019, 30(7): 875−877. DOI: 10.3969/j.issn.1003-6350.2019.07.018.

    CUI C K, LI J H, ZENG P C, et al. Clinical value of multi-slice spiral CT angiography in the diagnosis of coronary artery lesions[J]. Hainan Medical Journal, 2019, 30(7): 875−877. DOI: 10.3969/j.issn.1003-6350.2019.07.018. (in Chinese).
    [18]
    王磊. 多层螺旋CT门静脉血管成像对食管胃底静脉曲张出血的预测价值[J]. 实用医技杂志, 2018, 25(8): 861−863. DOI: 10.19522/j.cnki.1671-5098.2018.08.026.
  • Related Articles

    [1]LI Xiamin, HUANG Xiaoying, LI Zhifei, GUO Fangkai, WANG Pengchao, SHAN Chunhui. Application of Dual Source Mono+Combined with Cinematic Rendering in Cirrhosis Portal Vein Imaging[J]. CT Theory and Applications, 2025, 34(1): 44-50. DOI: 10.15953/j.ctta.2024.149
    [2]LI Wen, ZHANG Zhiwei, ZUO Ziyu, WAN Chengxin, TANG Yu, YANG Yajing. The Value of Virtual Monoenergetic Computed Tomography Angiography with Dual-layer Detector Spectral Computed Tomography for Imaging Cerebral Vessels[J]. CT Theory and Applications, 2024, 33(6): 669-675. DOI: 10.15953/j.ctta.2024.074
    [3]DENG Xiaolin, XIE Hui, QU Yalin, WU Shaoquan, ZHOU Bangjian, YANG Zhengbin, FENG Lian, WANG Min. The Value of Optimal Monochromatic Spectral CT Imaging in Gastric Variceal Bleeding before Endoscopic Treatment[J]. CT Theory and Applications, 2021, 30(5): 603-609. DOI: 10.15953/j.1004-4140.2021.30.05.08
    [4]KONG Lingling, XU Chijie, ZHAO Minyue, DENG Xiaoyi. The Application Value of Dual-Energy CT Virtual Monoenergetic Imaging in the Differential Diagnosis of acute and Chronic Spinal Compression Fractures[J]. CT Theory and Applications, 2021, 30(2): 209-216. DOI: 10.15953/j.1004-4140.2021.30.02.08
    [5]HAN Xinjun, ZHAO Liqin, AN Weimin, LIU Changchun, CAO Qiuting, SHANG Shufan. Study of the Association between Spleen-related Parameters and the Degree of Esophageal Varices Using Dual Energy CT[J]. CT Theory and Applications, 2019, 28(5): 593-600. DOI: 10.15953/j.1004-4140.2019.28.05.09
    [6]WANG Weixin, ZHANG Qiuhuan, GUO Pengde, HE Feifei, LIU Ming, LI Jie, CHEN Zhengguang. The Evaluation of Imaging Quality of Abdominal Vein by Combining Spectral CT Optimal Monotromatic Imging and ASiR Technique[J]. CT Theory and Applications, 2019, 28(1): 61-72. DOI: 10.15953/j.1004-4140.2019.28.01.07
    [7]HE Xiao-qing, ZHU Wan-shou, LIANG Han-huan, PENG Ke-yu. The Value of Dual-source CT Dual-energy Imaging in the Diagnosis of Gouty Arthritis[J]. CT Theory and Applications, 2018, 27(2): 171-177. DOI: 10.15953/j.1004-4140.2018.27.02.05
    [8]LU Yan-yan, CHEN Hong-wei, BAO Jian, FANG Xiang-ming, ZOU Xin-nong, TANG Qun-feng. Study on Clinical Application of Dual-source CT for Differentiating Diagnosis between Renal Clear Cell Carcinoma and Non-clear Cell Carcinoma[J]. CT Theory and Applications, 2018, 27(1): 19-25. DOI: 10.15953/j.1004-4140.2018.27.01.03
    [9]SHI Cen, ZHANG Huan, YAN Jing, LIU Huan-huan, PAN Zi-lai, YAN Fu-hua. Value of Advanced Virtual Monoenergetic Images in Evaluating Early Gastric Cancer[J]. CT Theory and Applications, 2014, 23(4): 535-540.
    [10]ZHAO Li-qin, HE Wen, LI Jian-ying, CHEN Jiang-hong, HU Zhi-hai, ZHANG Rui. The Pilot Study of Energy Spectral CT on the Image Quality of CT Portal Venography[J]. CT Theory and Applications, 2011, 20(3): 383-390.

Catalog

    Article views (48) PDF downloads (5) Cited by()
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return