医用CT仪上基于衰减的曝光量和峰值电压自动控制基本等式

ZHENG Xiao-ming

ZHENG Xiao-ming. 医用CT仪上基于衰减的曝光量和峰值电压自动控制基本等式[J]. CT理论与应用研究, 2016, 25(6): 625-632. DOI: 10.15953/j.1004-4140.2016.25.06.01
引用本文: ZHENG Xiao-ming. 医用CT仪上基于衰减的曝光量和峰值电压自动控制基本等式[J]. CT理论与应用研究, 2016, 25(6): 625-632. DOI: 10.15953/j.1004-4140.2016.25.06.01
ZHENG Xiao-ming. Attenuation Based Governing Equations for Automatic mAs and kVp Controls in Medical X-ray CT Imaging[J]. CT Theory and Applications, 2016, 25(6): 625-632. DOI: 10.15953/j.1004-4140.2016.25.06.01
Citation: ZHENG Xiao-ming. Attenuation Based Governing Equations for Automatic mAs and kVp Controls in Medical X-ray CT Imaging[J]. CT Theory and Applications, 2016, 25(6): 625-632. DOI: 10.15953/j.1004-4140.2016.25.06.01

医用CT仪上基于衰减的曝光量和峰值电压自动控制基本等式

详细信息
    作者简介:

    ZHENG Xiao-ming: Xiaoming Zheng is currently a senior lecturer at Charles Sturt University Australia teaching computerised tomography and magnetic resonance imaging. He was educated in China with B.Sci. and M.Sci. degrees from Zhejiang University. He earned his Ph.D. in Physics from the University of Newcastle Australia and subsequently worked as a postdoctoral fellow in applied mathematics at the University of Wollongong and an associate lecturer in theoretical chemistry at the University of Sydney. He was a medical physicist in nuclear medicine at the Prince of Wales Hospital Sydney prior to his current appointment at Charles Sturt University. He spent some 6 months on a special study program in the department of radiology at the University of Massachusetts Medical School, USA.

  • 中图分类号: TP391.41

Attenuation Based Governing Equations for Automatic mAs and kVp Controls in Medical X-ray CT Imaging

  • 摘要: 为了给患者以最少的辐射剂量而不影响诊断,现行医用CT仪都有基于噪声指标的自动曝光量控制。峰值电压对于图像质量和辐射剂量影响都很大,因此自动峰值电压控制很重要。本文从图像质量和辐射剂量优化的第一原理出发,给出基于衰减的曝光量和峰值电压同时自动控制的一般等式。此等式可用于单独曝光量自动控制或者单独峰值电压自动控制。对于单独暴光量自动控制,此等式要求对患者体厚度毎增加一厘米,曝光量要增加3.8%。对于单独峰值电压自动控制,此等式要求对患者体厚度每增加一厘米,峰值电压要平均增加1.53%。如果患者体厚度是常数,此等式要求平均百分比曝光量增加是百分比峰值电压减少的2.49倍。这些等式应该可以在现行医用CT仪上用于曝光量和峰值电压的自动控制。
    Abstract: Automatic exposure control based on noise index has been commonly implemented in current medical X-ray CT scanners in order to minimise radiation dose to patients while maintaining image quality for patient diagnosis. Automatic peak voltage control is also desirable as it is an important scanning parameter that has significant impact on both image quality and radiation dose. In this paper, a general attenuation based governing equation for simultaneous automatic controls of both mA s and kV p is presented which was derived from the first principle of radiation dose and image quality optimisation. This equation can be applied separately for automatic mA s control only or automatic kV p control only. The equation for automatic mA s control only states that the mA s should be increased by 3.8% for every 1cm increase of the patient body's sectional thickness if the peak voltage is kept constant. For automatic kV p control only, the equation states that the peak voltage should be increased by 1.53% percent on average for every 1 cm increase of the patient's body sizes if the mA s is kept as a constant. For a constant body thickness, the equation states that the average percentage mA s reduction should be 2.49 times the percentage peak voltage increase. The implementation of these equations in medical CT scanners has been described in a recent Australian patent.
  • 期刊类型引用(5)

    1. 刘佳萍,张卫国. 胸部定量CT对COPD患者治疗反应预测因素分析. 河北医药. 2024(03): 376-379 . 百度学术
    2. 曹亚娟,王警建,金晨望. 各期COPD支气管管壁和DLco的相关性分析. 贵州医药. 2024(06): 862-864 . 百度学术
    3. 马俊丽,舒建军,何平,苏海龙. Revolution低剂量CT扫描定量评估与慢性阻塞性肺疾病严重程度的相关性研究. 医学信息. 2023(14): 125-128 . 百度学术
    4. 郭建婷,龙怀聪. 慢性阻塞性肺疾病小气道的检测方法及研究进展. 国际呼吸杂志. 2022(14): 1109-1114 . 百度学术
    5. 王挺年. 定量CT参数评价COPD疾病严重程度的研究. 中国现代医生. 2022(31): 60-63 . 百度学术

    其他类型引用(1)

计量
  • 文章访问数:  584
  • HTML全文浏览量:  1
  • PDF下载量:  7
  • 被引次数: 6
出版历程
  • 收稿日期:  2016-06-11
  • 网络出版日期:  2022-12-04
  • 发布日期:  2016-12-24

目录

    /

    返回文章
    返回
    x 关闭 永久关闭