320排CT前瞻性心电门控容积扫描窄采集窗设置的临床评价

邢桂荣, 包丽丽, 刘挨师

邢桂荣, 包丽丽, 刘挨师. 320排CT前瞻性心电门控容积扫描窄采集窗设置的临床评价[J]. CT理论与应用研究, 2016, 25(4): 445-452. DOI: 10.15953/j.1004-4140.2016.25.04.08
引用本文: 邢桂荣, 包丽丽, 刘挨师. 320排CT前瞻性心电门控容积扫描窄采集窗设置的临床评价[J]. CT理论与应用研究, 2016, 25(4): 445-452. DOI: 10.15953/j.1004-4140.2016.25.04.08
XING Gui-rong, BAO Li-li, LIU Ai-shi. 320-Detector Row CT Narrow Acquisition Window in Single Cardiac Cycle for CTCA: Clinical Evaluation[J]. CT Theory and Applications, 2016, 25(4): 445-452. DOI: 10.15953/j.1004-4140.2016.25.04.08
Citation: XING Gui-rong, BAO Li-li, LIU Ai-shi. 320-Detector Row CT Narrow Acquisition Window in Single Cardiac Cycle for CTCA: Clinical Evaluation[J]. CT Theory and Applications, 2016, 25(4): 445-452. DOI: 10.15953/j.1004-4140.2016.25.04.08

320排CT前瞻性心电门控容积扫描窄采集窗设置的临床评价

基金项目: 

内蒙古自治区卫生和计划生育委员会科研项目(201301050)。

详细信息
    作者简介:

    邢桂荣(1965-),女,内蒙古医科大学附属医院影像诊断科主任护师,研究方向CT影像技术,Tel:15547120845,E-mail:duangsk@163.com;刘挨师*(1965-),男,内蒙古医科大学附属医院影像诊断科主任,教授,主任医师,研究方向CT血管成像,Tel:13948439219,E-mail:liuaishi@sina.com。

  • 中图分类号: R812;R445

320-Detector Row CT Narrow Acquisition Window in Single Cardiac Cycle for CTCA: Clinical Evaluation

  • 摘要: 目的:通过低心率下宽采集窗与窄采集窗图像质量与辐射剂量对比,明确窄采集窗320排CT冠脉造影的临床应用价值。方法:前瞻性选取2015年9月至2016年1月拟行CTCA且心率(HR)≤65次/min、心率变异度(HRv)≤5次/min的患者80例,平均年龄51岁(范围:36岁83岁)。所选患者随机分为A、B两组,其中A组(n=40)采集期相设置为R-R间期70%~80%。B组(n=40)采集期相设置为R-R间期30%~80%。两组设置管电压100 k V,依据体质量指数自动设置管电流300~580 mA。评价并比较A、B两组辐射剂量与图像质量。结果:A、B两组辐射剂量差异具有统计学意义(2.02 mSv±0.78 mSv vs 4.35 mSv±0.85 mSv,t=-12.474,P=0.000)。A和B两组可评估率差异无统计学意义(χ2=0.215,P=0.643)。A组平均评分3.61±0.53,B组平均评分3.71±0.48,Z=-3.074,P=0.002,差异有统计学意义。A、B两组主动脉根部衰减值差异无统计学意义(t=-1.132,P=0.261)。结论:320排螺旋CT窄采集窗设置可以提供可评估的图像质量,从而大幅度减少辐射剂量。
    Abstract: Objective: To evaluate the value of narrow acquisition acquisition CTCA by single cardiac cycle using 320-detector row CT. Methods: 80 patients(HR ≤ 65 bpm, HRv ≤ 5 bpm) were enrolled from september 2015 to January 2016 in this study. All of patients were divided into two groups. Group A included 40 patients with acquisition phase 70%~80% in R-R interval and Group B with acquisition phase 30%~80% in R-R interval. Both two groups with tube voltage 100 k V and tube current tailored to BMI(300~580 mA). Two groups were evaluated by radiation dose and image quality(all coronary artery segments scaleand percentage of assessable, aorta attenuation). A P value below 0.05 was considered to be statistically significant. Results: All patients include group A and B successfully underwent single cardiac cycle CTCA. The mean effective dose for group A was 2.02 mSv ± 0.78 mSv; this represents a 46% reduction as compared with that group B(4.35 mSv ± 0.85 mSv,t=-12.474,P=0.000). The mean image quality score was 3.61 ± 0.53 obtained with Group A versus 3.71 ± 0.48 obtained with Group B(Z=-3.074,P=0.002). The difference of the percentage of assessable coronary artery segments in both groups was not statistically significant(χ2=0.215,P=0.643). The difference of aorta attenuation in both groups was not statistically significant(t=-1.132,P=0.261). Conclusions: narrow acquisition acquisition CTCA may reduce the radiation dose keeping image quality assessable in patients with low HR(≤ 65 bpm) and HRv(≤ 5 bpm).
  • 期刊类型引用(5)

    1. 谭渊,胡春阳,陈钇地,向辉华,杨早,黎学兵. 128层CT肩关节不同扫描方案的图像质量与辐射剂量研究. 实用医学影像杂志. 2023(02): 97-100 . 百度学术
    2. 廖焕元,林盛义,程启彬. 等中心双低CT扫描联合迭代重建技术在肩关节创伤中的应用研究. 吉林医学. 2020(03): 530-532 . 百度学术
    3. 彭睿,方正,陈金华,倪卫国,石永贵,万千,黄飞. 缩窄320排CT冠状动脉成像采集时间窗的可行性. 中国医学影像技术. 2020(03): 460-463 . 百度学术
    4. 彭广宇,廖焕元,刘金丰. 双低CT扫描联合迭代重建在鼻骨骨折诊断中的应用. 中国数字医学. 2020(10): 125-127 . 百度学术
    5. 张卓璐,李少朕,洪楠,刘卓,陈雷. 采集时间窗对冠状动脉CT图像质量及辐射剂量的影响. 临床放射学杂志. 2019(02): 338-341 . 百度学术

    其他类型引用(1)

计量
  • 文章访问数:  474
  • HTML全文浏览量:  11
  • PDF下载量:  6
  • 被引次数: 6
出版历程
  • 收稿日期:  2016-02-21
  • 网络出版日期:  2022-12-04
  • 发布日期:  2016-08-24

目录

    /

    返回文章
    返回
    x 关闭 永久关闭